Parameter
Range
Whole blood
reproducibility
MFR ≥ 20%
HFR
RBC ≥ 3.00 × 10
12
/L RET%: 1.00% ~
4.00%
100% or ± 2.0%
IRF
RBC ≥ 3.00 × 10
12
/L RET%: 1.00% ~
4.00%
30%
IRF 20%
IPF
PLT 50×10
9
/L
IPF 3.0%
25%
C.5 Carryover
Parameter
Deviation of reading
WBC
1.0%
RBC
1.0%
HGB
1.0%
PLT
1.0%
D. Keyboard (optional)
USB port (supporting the protocol of USB2.0 and above) keyboard.
E. External Barcode Scanner (optional)
USB port (supporting the protocol of USB2.0 and above) hand-held barcode scanner.
F. Printer (optional)
HP Laser Jet Pro M404n and HP Office Jet Pro 8210
G. Interfaces
4 USB ports (3 USB 2.0 and 1 USB 3.0)
1 Ethernet port
H. Power Supply
Voltage
Input Power
Frequency
(100V240V~) ± 10%
300 VA
(50 Hz/60 Hz) ± 1 Hz
I. Fuse
WARNING
The fuse of the analyzer is not replaceable. In case of any questions, contact customer service
department.