Page 64
C. Performance specifications
C.1 Display range
Parameter
Display range
WBC
0.00 × 10
9
/L ~ 999.99 × 10
9
/L
RBC
0.00 × 10
12
/L ~ 99.99 × 10
12
/L
HGB
0 g/L ~ 350 g/L
PLT
0 × 10
9
/L ~ 9999 × 10
9
/L
MCV
0 fL ~ 450 fL
C.2 Background/blank count
Parameter
Background/blank count requirements
WBC
≤ 0.10 x 10
9
/L
RBC
≤ 0.02 x 10
9
/L
HGB
≤ 1 g/L
PLT
5 x 10
9
/L
C.3 Linearity and deviation range
Parameter
Linearity range
Deviation range
WBC
(0.00 100.00) x 10
9
/L
± 0.20 x 10
9
/L or ± 3%
(100.01 350.00) x 10
9
/L
± 6%
(350.01 500.00) x 10
9
/L
± 11%
RBC
(0.00 8.00) x 10
12
/L
± 0.03 x 10
12
/L or ± 2%
(8.01 16.99) x 10
12
/L
± 0.06 x 10
12
/L or ± 4%
HCT
(0 75) %
± 1.0 (HCT value) or ± 2%
HGB
(0 250) g/L
± 2 g/L or ± 2%
PLT
(0 1000) × 10
9
/L
± 10 x 10
9
/L or ± 5%
(1001 5000) x 10
9
/L
± 6%
RET#
(0.0000 0.8000) x 10
12
/L
± 0.0150 x 10
12
/L or ± 20%
RET%
(0.00 30.00%)
± 0.30 (RET% value) or ± 20%

Table of Contents