Vers.: 20230710ENG
Page 40
6.2.10 Flags of analysis results
The analyzer provides two types of flags for analysis results: parameter flags (1) and flags for
abnormal blood cell differential or morphology (2).
1. The analyzer provides the following parameter flags:
a. If the parameter is followed by an H or L, it means the analysis result has
exceeded the upper or lower limit of the reference range. See Section 9.2.4
Parameter Setup.
b. If the parameter is followed by an "R", it means the analysis result is
questionable.
c. If you see ****, as opposed to the result, it means the result is invalid; if you see
++++ as opposed to the result, it means the result is out of the display range.
NOTE:
The results of a background check will not be flagged for abnormal parameters, abnormal
blood cell differential or morphology.
2. The analyzer reports the flags for the following abnormal blood cell differential or
morphology.
High WBC analysis results
WBC# > 20% above
upper normal range limit.
Low neutrophils analysis results
NEUT# > 20% below
lower normal range limit.
High neutrophils analysis results
NEUT# > 20% above
upper normal range limit.
Low lymphocytes analysis results
LYMPH# > 25% below
lower normal range limit.
High lymphocytes analysis results
LYMPH# > 25% above
upper normal range limit.
High monocytes analysis results
MONO# > 40% above
upper normal range limit.
High eosinophils analysis results
EOS# > 40% above
upper normal range limit.
High basophils analysis results
BASO# > 100% above upper
normal range limit.
Low hematocrit analysis result
HCT > 10% below
lower normal range limit.
Polycythemia /
Erythrocytosis
High hematocrit analysis result
HCT > 10% above
upper normal range limit.
MCV > 10% below
lower normal range limit.
MCV > 10% above
upper normal range limit.