Vers.: 20230710ENG
Page 27
*Connection of the waste container (1: floater sensor and 2: tube), the DS Diluent (3), the Solution Reagent (4), the LH Lyse
(5), the LD Lyse (6) and the DR Diluent (7)
CAUTION
Make sure the diluent pipe and waste pipe are no longer than 1500 mm.
The waste container must be placed lower than the countertop that accommodates the
analyzer. The diluent container can be placed on the countertop or below the analyzer.
NOTE:
The analyzer performance may be undermined if it has been placed in an environment of
high dustiness.
The surface of the analyzer shall be cleaned and sterilized regularly with alcohol (75%).
The probe wipe block of the analyzer shall be wiped with alcohol (75%) regularly.
Sample collection and preparation must be done following standard procedures.
If any tubing or fluidic components is worn out, stop using the analyzer and contact the
technical support department immediately for inspection or replacement.
Check and make sure the reagents, lyse and waste tubing are not pressed or bent.
Note the expiration dates and open-container stability days of all the reagents. Do not use
expired reagents.