Vers.: 20230710ENG
Page 27
*Connection of the waste container (1: floater sensor and 2: tube), the DS Diluent (3), the Solution Reagent (4), the LH Lyse
(5), the LD Lyse (6) and the DR Diluent (7)
CAUTION
• Make sure the diluent pipe and waste pipe are no longer than 1500 mm.
• The waste container must be placed lower than the countertop that accommodates the
analyzer. The diluent container can be placed on the countertop or below the analyzer.
NOTE:
• The analyzer performance may be undermined if it has been placed in an environment of
high dustiness.
• The surface of the analyzer shall be cleaned and sterilized regularly with alcohol (75%).
• The probe wipe block of the analyzer shall be wiped with alcohol (75%) regularly.
• Sample collection and preparation must be done following standard procedures.
• If any tubing or fluidic components is worn out, stop using the analyzer and contact the
technical support department immediately for inspection or replacement.
• Check and make sure the reagents, lyse and waste tubing are not pressed or bent.
• Note the expiration dates and open-container stability days of all the reagents. Do not use
expired reagents.