3.7 Connection with the System Controller
3.7.1
Setting the Instrument
To control the instrument from the system controller CBM-40 CL set the parameters as
follows:
Setting Parameter Set Value Reference
LOCAL MODE 0 : Remote
"Selecting the local mode
LOCAL MODE" P.24
LINK ADDRESS Link address
"Setting the link addressLINK
ADDRESS" P.24
OPERATION MODE 0 : 40C
Not used in CL model.
Do not change the factory default value.
3.7.2
Basic Parameters
CBM-40 CL can connect to a maximum of four column ovens. The system controllers
configure the settings such as the operating temperature, upper temperature limit,
automatic column switch valves 1 and 2, ON/OFF of temperature control, and time
program. Additionally, the workstations support column management functions. For
details, refer to the instruction manuals of the system controllers and workstations.