5.1 Specifications
The performance data given below is based at room temperature of 25 °C.
Item Specification
Type Air circulation
Temperature control
range
(Room temperature* - 10 °C) to (100 °C)
* Room temperature means the ambient temperature at the CTO-40C CL.
The upper limit is 85 °C when an automatic column
switch valve or column management device (CMD) is
connected.
Temperature setting
range
4 °C to 100 °C in 1 °C steps
Temperature control
precision
0.1 °C (± 0.05 °C)
Ambient temperature
range
4 °C to 35 °C
Size and number of
columns stored
Six columns up to 250 mm or three 250 mm to 300 mm columns
Safety measures
• Upper temperature limit can be set to prevent inadvertent
accidents.
• Equipped with the thermal fuse to prevent overheating at
breakage
•
Incorporating the gas sensor
• Incorporating the liquid leak sensor
Time program
functions
Change the temperature setting (in steps or linearly), start or stop
temperature control, change the flow lines of automatic column
switch valves, and repeat a program, 100 steps, 0.01 to 999.99 min
Dimensions
W 250 mm ´ H 415 mm ´ D 500 mm (Excluding a protrusion)
Mass 21 kg
5
Technical Information
CTO-40C CL
61