4.1.1
Prior to Inspection and Maintenance
Replace the mobile phase solvent in the flow lines with water.
Wipe away any dirt from the left and right doors and the main cover.
Wipe away any dirt from the control pannel with tissue paper or a soft cloth moistened
with water.
4.1.2
List of Periodic Inspection and Maintenance
The maintenance and replacement periods listed in this table are presented only
as guidelines. These are not guarantee periods.
These will vary depending on usage conditions.
Inspection/
Maintenance Item
1 Year 2 Years 3 Years Remark Ref.
Air filter replacement
P
      P.58
Inspection of Temperature
Accuracy
P
Nexera
CL
seires
System
Guid e
"4.8.2
Testing
the
Temper
ature
Adjust
ment
Function"
Inspection of Gas Sensor
P
P.37
Inspection of Leak Sensor
P
P.36
CTO-40C CL
55
4.1 Periodic Inspection and Maintenance
4

Table of Contents