P., 'Cil 'NJ 14
'IH 'LL
',A_
•
• •
I @ ir
4 INSTALLATION AND MAINTENANCE/INSPECTION
MUX-100 SERVICE MANUAL 4 - 14
4-2-8 User Setting
According to the request from the customer, change each user setting.
Ⅰ
How to enter user setting mode
Press and hold t
he user setting switch for 3 seconds to change into user setting mode. .
“usroPe” is displayed (for approximately 1 second), then the contents of user setting are displayed in the tube
voltage display unit. The contents of user setting can be changed by using the tube voltage setting keys “△”
and “▽”.
kV display Item
PAS Power-assist setting
Acc Acceleration setting
StC Steer control setting
bAC Backward power-assist setting
APO Auto-power-off time setting
bON Buzzer control setting
LAP Collimator lamp control setting
aPr APR setting at the time of starting the apparatus
id Apparatus ID setting
Table 4-6 Contents of User Setting
Ⅱ
How to return to normal use status
Press the user setting switch in user setting mode.
At this time, if the contents of user setting are not registered by the registration key, the changing of contents are
canceled.
①
Press and hold the user setting switc
for 3 seconds.
③ ②
The system enters the user setting mode such as PAS, Acc,
StC etc. (The photograph on the left shows the power assis
setting mode PAS.)
Change each set item using the tube voltage setting keys, an
register each set item using the registration key. When a se
item is re
istered
the s
stem returns to the normal use status.
Press the user setting switch t
return to the normal use status.
④