4 INSTALLATION AND MAINTENANCE/INSPECTION
MUX-100 SERVICE MANUAL 4 - 26
Tube arm Any scratches or abnormal sound occurring during
up/down movement?
Bumper Do you operate normally?
Any loose parts?
Earth belt Does the belt touch the floor?
The belt is not short or cut?
(Copper wire in the lower portion of the main
unit)
Remove dust,
when dust has
adhered to the
belt.
DC motor The key on the motor shaft is not loose?
The wheel doesn’t shake?
Brake for DC motor The bar of a travel brake operate certainly?
Casters
The fixing of each casters are not loose?
(4 mounting screws provided in the lower portion of the
main unit)
FPD Bin Are support cushions obstructed?
Grid holder Is not the part where the grid case is installed
transformed?
Cable Any twisting or damage to the cable?
Internal inspection Confirm the connectors.
Remove the
side covers
Arm covers
Top cover
Power plug Is there any gap between the plug and the stopper?
Are any of the plug’s prongs bent or deformed?
Cord reel Is the cable cover cut or frayed?
Is anything caught in it and making noise?
Table 4-10 Periodical check items