7.3 Measuring parameters
Variant Components
DeSO
x
SO
2
, CO
2
, optional O
2
Full configuration SO
2
, CO
2
, NO, NO
2
, optional O
2
Number of measured variables
Number of measured variables Max. 5
Measuring method
Measuring method Cold-extractive
Sample volume
Sample volume 60 ... 100 l/h
Spectral range
Spectral range UV, VIS
Component Measuring ranges
SO
2
0 ... 100 ppm; 0 ... 500 ppm
CO
2
0 ... 25% by volume
O
2
0 ... 21%volume
NO 0 ... 300 ppm; 0 ... 1500 ppm
NO
2
0 ... 200 ppm; 0 ... 500 ppm
Measuring point switchover
Measuring point switchover Max. 4 measuring points
Measured value characteristics
Measuring precision < 1% of the respective full scale value
Detection limit < 0.5% of the respective full scale value
Sensitivity drift < 2% of the respective full scale value per week
Zero drift < 2% of the respective full scale value per week
Span drift < 2% of the respective full scale value per week
Setting time t
90
15 ... 30 s, total active measuring path as from sampling
(With advance extraction with several measuring points)
7.4 Ambient conditions
Ambient conditions in operation
Installation location Below deck
Ambient temperature +5 ... +45 °C
Relative humidity < 90% (without condensate)
Air pressure 900 ... 1100 hPa
Degree of protection IP 54
TECHNICAL DATA 7
8017324/15A2/V6-0/2019-10 | SICK T E C H N I C A L I N F O R M A T I O N | MARSIC200
61
Subject to change without notice