Wait time after switching to sample gas, then the manual adjustment is
regarded as completed
ä
Start the selected manual adjustment
1
Empty field: No suitable test gas setting programmed for the selected function
2
Actual value for the adjustment = mean value of measured values within the measuring time
Result
Figure 23: Menu Manual adjust with result (example)
1
Cancel the running manual adjust
2
Error shown when no result available. Erroneous sequence or drift overrun.
3
The actual measured value measured
5.3 Automatic adjustments/validations
5 ADJUSTMENT FUNCTIONS
44
T E C H N I C A L I N F O R M A T I O N | MARSIC200 8017324/15A2/V6-0/2019-10 | SICK