Power input, complete system (max. for each 5 m sample gas line)
4 4000 VA 17 A 35 A
Connections
Connections, analyzer
•
Signal cable, 7-wire, 1.5 mm
2,
shielded, e.g., Marine‐
line (Sick Part No. 6056230)
•
Power cable, 3-wire, min. 1.5 mm², e.g., Marineline
(Sick Part No. 6056215)
Distribution unit
•
Signal cable, 7-wire, 1.5 mm
2,
shielded, e.g., Marine‐
line (Sick Part No. 6056230)
•
Signal cable, 4-wire, 0.75 mm
2,
shielded, e.g.,
Marineline (Sick Part No. 6056229)
•
Power cable, 3-wire, min. 1.5 mm², e.g., Marineline
(Sick Part No. 6056215)
Sample conditioning
•
Signal cable, 4-wire, 0.75 mm
2,
shielded, e.g.,
Marineline (Sick Part No. 6056229)
•
Power cable, 3-wire, min. 1.5 mm², e.g., Marineline
(Sick Part No. 6056215)
Sample gas line
•
Power cable, 3-wire, 1.5 mm²
Sampling probe
•
Signal cable, 4-wire, 0.75 mm
2,
shielded, e.g.,
Marineline (Sick Part No. 6056229)
•
Power cable, 3-wire, min. 1.5 mm², e.g., Marineline
(Sick Part No. 6056215)
7.10 Emissions
Emissions
Condensate For water vapour-saturated exhaust gas (e.g., after
exhaust gas washer): Approx. 2 liters of condensate per
week
7.11 Torques for screw fittings
A medium total friction coefficient of µ=0.12 is assumed in the specification of torques.
This means that the screws are installed lightly oiled.
Analyzer Property class Torque
Analyzer door screw fitting A2-50 3 Nm
Analyzer enclosure 4 screws M8
bolts
and bracket on analyzer enclosure
8.8 23.1 Nm
10.9 34 Nm
12.9 39.6 Nm
A2/4-50 7.1 Nm
A2/4-70 16 Nm
A2/4-80:9 22 Nm
FINOR 3 screws base plate 4 Nm
Ground screw FINOR 3 Nm
Ground screw analyzer 3 Nm
Gas module 4 screws base plate 3 Nm
Complete analyzer module, 7 black
screws
9 Nm
TECHNICAL DATA 7
8017324/15A2/V6-0/2019-10 | SICK T E C H N I C A L I N F O R M A T I O N | MARSIC200
65
Subject to change without notice