EasyManuals Logo

SICK MARSIC200 User Manual

SICK MARSIC200
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #63 background imageLoading...
Page #63 background image
7.7 Sample gas line, heated
Sample gas line
Ambient temperature –20 ... 80 °C
Working temperature +120°C
Heating Self-regulating parallel heater line
Power supply 115 V / 230 V
Power consumption 60 W/m at 10 °C
Dimensions see "Sample gas line, heated", page 58
Degree of protection IP 54
Protection class I
7.8 Interfaces and protocols
Operation and interfaces
Operation Via LC-Display or SOPAS ET software, several operating
levels, password-protected
Display and input Black-and-white foiled screen with function buttons
Status LEDs:
“Power”
“Malfunction”
“Maintenance request”
Remote control Ethernet (TCP/IP):
Connector: RJ 45
Type: TCP/IP Peer-to-Peer.
Method: 10 MBit half-duplex
Modbus
Remote maintenance SICK MPR (optional)
PC operation SOPAS ET via Ethernet
Analog outputs
Number 8
Reference potential Potential-free (electrically isolated)
Signal range 0 … 24 mA
Residual ripple 0.02 mA
Resolution/precision 0.1% (20 μA)
Accuracy 0.25% of full-scale value
Maximum load 500 Ω
Maximum output voltage 15 V
Start or error state Adjustable
Analog inputs
Number 2
Reference potential GND
Input signal 0 … 20 mA
Highest allowable input signal 30 mA
Overcurrent protective device ±1000 mA
TECHNICAL DATA 7
8017324/15A2/V6-0/2019-10 | SICK T E C H N I C A L I N F O R M A T I O N | MARSIC200
63
Subject to change without notice