5.4.10 Connecting the exhaust gas outlet
Important information
WARNING
Noxious and aggressive exhaust gases
The exhaust gases may contain harmful or irritating components.
b
Lead the measuring system gas outlets outdoors or into a suitable flue.
b
Do not connect the exhaust gas line with the exhaust gas line of sensitive subas‐
semblies. Aggressive gases could damage these subassemblies as a result of
diffusions.
NOTICE
Condensate could accrue in the exhaust gas line.
b
Use a suitable hose line to run the condensate outlet into an open condensate
container or a waste disposal line.
b
Lay the line so that it always runs downwards.
b
Keep the line opening free from any blockages or liquids.
b
Protect the line from frost.
Procedure
1. Connect the exhaust gas outlet at the intended place.
2. Lay the exhaust gas line in a suitable manner:
°
The gas outlet must be open to the ambient pressure; in waste disposal lines
it can be laid with a light partial vacuum.
°
Do not bend or crimp exhaust gas lines.
5 MOUNTING
30
O P E R A T I N G I N S T R U C T I O N S | MCS200HW 8021889/1D1T/V3-1/2021-09 | SICK