Component Measuring range
CO
2
0 ... 25% by volume; 0 ... 50% by volume
SO
2
0 ... 26 ppm; 0 ... 875 ppm
H
2
O 0 ... 40% by volume
O
2
0 ... 25% by volume
N
2
O 0 ... 50 ppm; 0 ... 1.015 ppm
TOC 0 ... 15 mg/m
3
; 0 ... 10,000 mg/m
3
Table 11: Certified measuring ranges in accordance with EN15267-3
Component Certified measuring ranges Additional measuring ranges
HCl 0 ... 15 mg/m
3
0 ... 3,000 mg/m
3
NH
3
0 ... 10 mg/m
3
0 ... 500 mg/m
3
CO 0 ... 75 mg/m
3
0 ... 10,000 mg/m
3
NO 0 ... 150 mg/m
3
0 ... 2,500 mg/m
3
CH
4
0 ... 50 mg/m
3
0 ... 500 mg/m
3
NO
2
0 ... 50 mg/m
3
0 ... 500 mg/m
3
SO
2
0 ... 75 mg/m
3
0 ... 2,500 mg/m
3
N
2
O 0 ... 100 mg/m
3
0 ... 2,000 mg/m
3
CO
2
0 ... 25% by volume —
H
2
O 0 ... 40% by volume —
O
2
0 ... 25% by volume —
TOC 0 ... 15 mg/m
3
0 ... 50/150/500 mg/m
3
Table 12: Measured value characteristics
Measured value characteristics
Measuring principle Photometric
Measuring precision < 2% of the respective full scale value
Detection limit < 2% of the respective full scale value
Sensitivity drift < 2% of the respective full scale value per week
Zero point drift < 2% of the respective full scale value per week
Span drift < 2% of the respective full scale value per week
Setting time t
90
< 200 s, total measuring path as from probe extraction
13.2.2 Ambient conditions
Table 13: Operation
Ambient conditions in operation
Installation location Indoor installation
Ambient temperature +5 ... +40 °C
Ambient temperature with cooling
unit option
+5 ... +50 °C
Relative humidity < 90% (without condensate)
Air pressure 850 ... 1100 hPa
Table 14: Storage
Ambient conditions in storage
Ambient temperature -20 ... +70 °C
Relative humidity < 90% (without condensate)
TECHNICAL DATA 13
8021889/1D1T/V3-1/2021-09 | SICK O P E R A T I N G I N S T R U C T I O N S | MCS200HW
63