9 Menus
9.1 Password
Configuration is only possible on level “Authorized Client”. Login is performed using tile
“Login” and a password prompt.
Password for “Authorized Client”: HIDE (preset)
9.2 Menu tree
Menu level Explanation
1 Tasks Quick access to the most important
functions for the operator
2 Adjustment
2.1 Adjustment
2.1.1 Zero point adjustment The measured value zero points are
adjusted while instrument air is fed.
2.1.2 Adjustment with internal adjustment filter The concentrations of measuring com‐
ponents are adjusted with an adjust‐
ment filter.
2.1.3 Reference point adjustment The concentrations of measuring com‐
ponents are adjusted while span gas is
fed.
2.1.4 O2 adjustment The zero and reference point is
adjusted while instrument air is fed.
2.1.5 Pressure adjustment Adjustment of pressure sensors.
2.2 Validation
2.2.1 Zero point validation The measured value zero points are
checked while instrument air is fed, but
not adjusted.
2.2.2 Validation with internal adjustment filter The concentrations of measuring com‐
ponents are adjusted with an adjust‐
ment filter, but not adjusted.
2.2.3 Reference point validation The concentrations of measuring com‐
ponents are checked while span gas is
fed, but not adjusted.
2.3 Span gas feed Different reference materials can be
controlled. No adjustment or validation
is performed.
2.4 Results
2.4.1 Adjustment factors Displays the adjustment factors for
span gas and adjustment with internal
adjustment filter.
2.4.2 Zero point drift Displays the determined percentage
deviation after zero point validation.