7 Commissioning
7.1 Prerequisites for switching on
Procedure
1. Check the measuring device.
2. Instrument air must be connected and open.
3. If the instrument air has changed: Check the instrument air quality.
4. Check pressure settings on the pressure reducer unit.
Related topics
Measuring device check: see "Checking the system", page 48
Instrument air quality: see "Gas supply", page 65
Pressure reducer unit setting: see "Setting the pressure reducer module", page 27
7.2 Switching on
Procedure
1. Switch on the external power disconnection unit.
SOPASair loading screen is displayed.
A countdown is shown on the display, counting down from 80.
The start screen opens. Display: System initialization
The measuring device heats up:Display: System heats.The status indicator is
orange. Heating process can take up to 2 hours.
Display: Premeasure. The status indicator is orange.
The status indicator is green. Display: Measure.The measuring device is ready for
operation.
2. When the yellow or red status indicator is on: Display logbook and clear error.
The measuring device is in operation.
Related topics
Error list: see "Error messages and possible causes", page 52
7.3 Recognizing the safe operating state
The system is in proper operation when:
A system check has been carried out according to the Maintenance plan before
commissioning and in running operation.
Only the green status indicator is on and Measuring is shown in the status bar.
When the yellow or red status indicator is on: Display logbook and clear error.
Related topics
Checking the system: see "Checking the system", page 48
Error list: see "Error messages and possible causes", page 52
7.4 Adjusting
7.4.1 Performing zero point adjustment
Overview
Menu: Tasks Zero point adjustment
As standard, the zero point adjustment is used to adjust the zero points of the meas‐
ured values while instrument air is fed.
7 COMMISSIONING
34
O P E R A T I N G I N S T R U C T I O N S | MCS200HW 8021889/1D1T/V3-1/2021-09 | SICK
Subject to change without notice

Table of Contents

Other manuals for SICK MCS200HW

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK MCS200HW and is the answer not in the manual?

SICK MCS200HW Specifications

General IconGeneral
BrandSICK
ModelMCS200HW
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals