EasyManuals Logo

Siemens Acuson S3000 User Manual

Siemens Acuson S3000
518 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #75 background imageLoading...
Page #75 background image
2 Safety and Care
Instructions for Use 2 - 45
Approved List of Disinfectants
Note: CIDEX OPA and Gigasept FF may discolor the transducer. There is no associated degradation
of imaging performance or transducer reliability. You can use Gigasept FF as a cleaner and a
disinfectant.
CIDEX
CIDEXPLUS
CIDEX OPA
Gigasept FF
Milton
PI-Spray II
Rely+On Virkon
Revital Ox
RESERT HLD
Sani-Cloth AF3
Sani-Cloth HB
Super
Sani-Cloth
4C1
          
6C1 HD
      
NA
 
NA
6C2
          
7CF1
       
NA
 
7CF2
       
NA
 
8C3 HD







NA


NA
EV-8C4
       
NA NA NA
EC9-4
      
NA
  
MC9-4
       
NA
 
9EVF4








NA


9L4
  
NA NA

NA NA
  
12L4
  
NA NA
  
NA NA NA
14L5
      
NA
  
14L5 SP
      
NA
  
18L6 HD
      
NA
 
NA
4P1
      
NA
  
4V1
          
4V1c
          
8V3











10V4
      
NA
  
V5Ms
     
NA
NA
NA
 
V7M
      
NA NA NA NA
CW2
      
NA NA
 
CW5
      
NA NA
 
 = approved
NA = not approved