Appendix F  Acoustic Output Reference 
Instructions for Use  F  - 3  
Display Accuracy and Measurement Uncertainty 
When a transducer is capable of exceeding a mechanical index (MI) or thermal index (TI) of 
1.0, the ultrasound imaging system displays indices starting from 0.4 in increments of 0.1 for all 
displayed values. 
A number of factors contribute to the estimation error for the displayed index. Variations among 
the following factors contribute to the index display error: transducers and systems, 
assumptions and approximations for real-time index calculations, and measurement 
uncertainty. The MI and TI display values should be interpreted as relative information and only 
used to assist the operators to implement the ALARA (As Low As Reasonably Achievable) 
principle for patient examinations; the acoustic output should be the lowest level required to 
satisfactorily perform the examination. 
Measurement uncertainty for acoustic pressure, acoustic intensity, power, and center frequency 
is within 12%, 25%, 10% and 5% respectively assessed by considering both the systematic and 
random uncertainties. The total estimated display accuracy is ±25% for MI and ±50% for TI. 
Definitions for these parameters can be found in the IEC 62359, "Ultrasonics – Field 
Characterization – Test Methods for the Determination of Thermal and Mechanical Indices 
Related to Medical Diagnostic Ultrasonic Fields."