5  Transducer Accessories and Biopsy 
5 - 12  Instructions for Use 
Needle Path Verification 
Before performing any patient procedure using a needle guide, you must verify that the path of 
the needle is accurately indicated by the on-screen guidelines. 
The needle guide is ready for patient use only after the path of the needle has been verified. 
Checklist of Items Required for Needle Path Verification: 
  Transducer with attached needle guide 
  Water-based coupling agent (gel) 
  Sterile transducer sheath 
  New, straight, biopsy needle 
  Sterilized container of sterilized and degassed water 
To verify the path of the needle: 
  WARNING: The system displays biopsy guidelines to indicate an anticipated needle path. The 
guidelines do not account for a bent or deflected needle and are not intended as an absolute 
reference. It is the user's responsibility to verify correct positioning of the needle during a biopsy 
or puncture procedure. 
  WARNING: Do not use a needle guide if the path of the needle is not accurately indicated by the 
on-screen guidelines. The path of the needle must display within the guideline. Contact your 
Siemens service representative if the needle path is not accurately indicated. 
1.  Attach the needle guide to the transducer. 
2.  Connect the transducer to the system and activate the transducer. 
3.  Set the system to the depth of the intended puncture procedure. 
4.   Activate the biopsy function to display the guidelines on the image screen.  
●  Select Biopsy and then select A or B.  
○  For systems without a touch screen, select Biopsy A or Biopsy B on the image 
menu.  
5.  Ensure the selected angle (Biopsy A or Biopsy B) corresponds to the needle guide 
attached to the transducer. 
6.  Immerse the head of the transducer into the degassed water and insert the needle into the 
needle guide. 
7.  Verify that the path of the needle displays according to the guidelines shown on the image 
screen. 
After verification, the needle guide is ready for use.