5  Transducer Accessories and Biopsy 
5 - 6  I nstructions for Use 
Gel Pad 
The gel pad is a disposable bacteriostatic standoff. It is used when superficial imaging requires 
an appropriate standoff for utilizing the focal zone of the transducer. The gel pad provides a 
fixed distance between the transducer face and the body surface. 
  WARNING: Ultrasound energy is transmitted more efficiently through the gel pad than through 
tissue. When using a standoff device of any kind (for example a waterpath or gel pad), the actual 
mechanical and thermal indices, MI and/or TI, may be higher than indicated in the output display 
on the system.  
Preparation for Use 
Before use, examine the gel pad for any material flaws. Any product showing flaws should not 
be used. 
Needle Guide Brackets 
  WARNING: Percutaneous procedures always involve heightened risk to the patient and to the 
operator handling biopsy needle guides. Clinicians using Siemens recommended biopsy devices 
under ultrasound guidance should be trained and must observe proper needle insertion 
sequencing with the needle guide in order to avoid undue discomfort and unnecessary risk and 
injury to the patient. 
Refer to the in-box instructions for attachment and care procedures for all needle guide 
accessories except as noted below. 
See also: EC-1 Needle Guide Bracket Kit, p. 5-7