3  System Setup 
3 - 6  Instructions for Use 
During the Move 
  Caution: When moving the ultrasound system, protect it from environmental changes including: 
moisture, winds, dirt and dust, and extreme heat or cold exposure. 
  Caution: Avoid moving the ultrasound system on outside surfaces with loose dirt, contaminates, 
or standing liquids. 
  Caution: Care should be taken to minimize shock and vibration of the ultrasound system. Avoid 
uneven surfaces that contain an abrupt height change or jarring surface irregularities. 
You can move the ultrasound system from room to room within a facility and easily reposition 
the system during an examination. Be careful on inclines and uneven surfaces. The ultrasound 
system can be moved across pavement and other hardened parking lot surfaces. 
Note: The wheels of the ultrasound system must be locked when transporting by vehicle. The 
ultrasound system must be sufficiently anchored to the vehicle floor or walls such that it does not shift 
or move during transport. 
After the Move 
  Caution: Make sure the ultrasound system has proper ventilation during operation. Do not 
position the system against walls or hard surfaces that would impede free ventilation around the 
system. 
  Caution: Do not allow linens, bedding, and/or hanging curtain partitions to block the ultrasound 
system's ventilation. 
  Caution: Obstructed fans can cause potential system overheating, system performance 
degradation, or failure. 
  Caution: Brakes are most effective on a level surface. Never park the system on an incline 
greater than five degrees. 
  Position system: Make sure the system is not placed against walls or fabrics that obstruct 
perimeter air flow to the system cooling fans. 
  Set brakes: Set the front and rear brakes. 
  Unlock the flat panel monitor. 
  Plug in cord: Plug the power cord into a hospital-grade or local equivalent wall outlet. 
If applicable, also plug the Ethernet cable into an appropriate connector. 
  Power ON: Power on ( ) the ultrasound system. 
  Check display: After the boot-up sequence is complete, verify that the image display is 
stable, that you can select a transducer, and that selections made on the control panel 
respond to your selection.