# Text Description Action
676 ! Database: delay limit ex‐
ceeded on stream %3
Temperature or pressure wait set in the
method has been exceeded.
Check temp or press controller.
Check wait_delta and maxwait in tem‐
perature or pressure controller table.
677 ? Next Stream Error Cannot locate next stream in sequence
after a stream step.
Check sequence to ensure that there is
an enabled stream.
678 ! End of Cycle occurred be‐
fore events completed
System error Contact Customer Support.
679 ! Application is out of service Attempt was made to move (approve) re‐
sults for an out-of-service application.
Put application in service.
680 ? Print job failed: %3 for Print‐
er: %4
Print failure. Check printer.
681 ! Application is disabled Attempt was made to set disabled appli‐
cation to run.
Enable application.
682 ! Database: I/O failure: %3 Bad status on AO, DO write. Check I/O channels.
683 ? Database: no normal se‐
quence for application: %3
No active sequence. Check sequences.
684 ? Database: no enabled en‐
tries in sequence
Cannot find enabled entry in sequence. Check sequence.
685 ? Printer: TCP connection or
queue failed
Printer connection failed. Check address in printer table.
686 ? Printer: TCP Print failed Communication failure with TCP printer. Try printing again.
687 ! Results for cycle %3 lost
due to SNE reset
This alarm marks results as uncertain un‐
til a cycle has been completed after re‐
setting unit.
Automatic repair: No action necessary.
688 ! Method is corrupt: %3 re‐
load from EZChrom
Method is corrupt. Download method from EZChrom.
689 ! Database: Fault on Slave
Application %3
A fault is being transferred from the slave
to the master to invalidate the master's
results.
Check fault in slave application.
690 ! Database: Slave Applica‐
tion not Complete %3
Master application is trying to run when
slave is not in hold. The slave must com‐
plete before the master.
Check cycle length of slave application;
should be shorter than master. Also
could occur if autocalibration sequence
for master is shorter than for the slave.
691 ? Database: Warning: %3 System error Contact Customer Support.
692 ! Database: Divide by zero in
%3
Peak measured value is zero during cal‐
ibration for an autocalibration, so margin-
checking cannot occur.
Check method.
693 ? Database: I/O warning: %3 System error Contact Customer Support.
694 ? Database:value > limit: %3 Limit exceeded. Message should contain
sufficient information.
See the Limits and Alarm Handlers ta‐
bles under the GCP Application View.
695 ? Database value < limit: %3 Limit exceeded. Message should contain
sufficient information.
See the Limits and Alarm Handlers ta‐
bles under the GCP Application View.
696 ? DB: Screen access denied System error Contact Customer Support.
697 ! DB: Run requested on dis‐
abled program: %3
System error Contact Customer Support.
General Maintenance and Troubleshooting
4.6 Alarm Codes, Descriptions, and Suggested Actions
Maxum edition II Analyzer General Maintenance
42 Manual, August 2018, A5E42019842001

Table of Contents

Other manuals for Siemens Maxum II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Maxum II and is the answer not in the manual?

Siemens Maxum II Specifications

General IconGeneral
CategoryGas Chromatograph
Measurement PrincipleGas Chromatography
Communication ProtocolsModbus, Ethernet, Profibus
Operating Temperature-20°C to 55°C
Power Supply24 VDC or 100-240 VAC
Mountingrack mount
ApplicationsRefinery, chemical, petrochemical, natural gas
DetectorsThermal Conductivity Detector (TCD), Flame Ionization Detector (FID)
Sample InletsMultiple
DimensionsVaries by configuration (e.g., 19" rack mount)

Related product manuals