# Text Description Action
3163,
3164
! EPC Shutdown on %4,
Channel [1 - 2]
The [n] channel of the Electronic Pres‐
sure Control (EPC) was temporarily shut
down.  This happens when the EPC
control valve is fully opened for more
than 5s and is meant to prevent over‐
heating of the valve.  This situation typ‐
ically happens when the input pressure
is too low, as from an empty cylinder.
No action required. The shutdown  is
temporary and the operation will auto‐
matically resume after 15s.
3203,
3204
! EPC ramp has no origin on
%4, Channel [1 - 2]
The starting point of the pressure ramp
was set to the current measured pres‐
sure as the setpoint value was not previ‐
ously set.   When no prior setpoint is
available, this alarm is issued and the
current measured pressure is used as a
substitute for setpoint #1.
Set a static setpoint first, and then set
the ramp rate, followed by a new set‐
point.  With this sequence, the starting
point of the ramp will be setpoint #1 and
the ending point will be setpoint #2.
4.6.16 Alarms 3401 - 3454 TFTP
GCP Alarm Descriptions 3401 - 3454 TFTP
The following tables list the alarm number (#), type (+ information, ? warning, ! error) alarm
text, description, and actions.
# Text Description Action
3401 ? TFTP Protocol Error Protocol error detected, such as receipt
of a non-DATA packet or lack of an ex‐
pected message acknowledgment. Most
likely the gateway settings on the SNE
are wrong.
Power up the SNE connected to a serial
terminal and change the settings in the
startup dialog.
3402 ? TFTP Timeout The TFTP client didn't receive a re‐
sponse from the server.
Verify that the TFTP server is running
on the host computer, and that the host
computer is connected to the SNE
through the network.
3403 ? TFTP Server out of Sync The data packets requested by the TFTP
client to not match those sent by the
TFTP server. Either the TFTP client or
TFTP server is not working properly.
Reset the SNE.
Restart the TFTP server on the host
computer.
3404 ? TFTP Server out of Sockets TFTP server cannot create a portal from
which to communicate. SNE or host com‐
puter is low on resources.
Wait for problem to clear.
Contact Customer Support.
3405 ? TFTP Max Channels ex‐
ceeded
Too many TFTP load requests have
been commanded.
Only request one TFTP load at a time.
3406 ? TFTP Driver Not Initialized TFTP driver was not successfully initial‐
ized. Usually a problem of resource.
Verify that the SNE has sufficient mem‐
ory.
Upgrade SNE.
Reset SNE.
3450 ? TFTP Client Out of Memory Client ran out of memory loading file. Verify that the SNE has 16Mb SIMM.
Reset SNE.
General Maintenance and Troubleshooting
4.6 Alarm Codes, Descriptions, and Suggested Actions
Maxum edition II Analyzer General Maintenance
70 Manual, August 2018, A5E42019842001