# Text Description Action
3757 ? NACK address on %4 SNECON PICs: An I2C message trans‐
action failed after the required number of
retries.  A cause of this failure was a
unacknowledged destination address
byte (first byte in the message).  This
may happen if a module was disconnec‐
ted from the bus after its address has
been reassigned.
Reset the analyzer.
3758 ? NACK Byte on %4 SNECON PICs: An I2C message trans‐
action failed after the required number of
retries. A cause of this failure was a un‐
acknowledged byte that is not the first
byte in the message (any byte other than
the destination address was not acknowl‐
edged). This may happen if a module is
seriously corrupted or the I2C signal in‐
tegrity is a problem (noise, etc…).
Contact Customer Support.
3759 ? NACK Message on %4 SNECON PICs: An I2C message trans‐
action failed after the required number of
retries.  A cause of this failure was a
NACK message with a non-zero flag.  
This may happen if the firmware of the
peripheral module is not responding
properly.Reset the analyzer.
Replace the module being addressed.
3760 ? Invalid Checksum on %4 SNECON PICs: An I2C message trans‐
action failed after the required number of
retries.  A cause of this failure was an
invalid checksum in the reply message.
 This may happen if the firmware of the
peripheral module is not working proper‐
ly or the electrical properties of the I2C
signals are marginal.
Contact Customer Support.
3761 ? Invalid Opcode on %4 SNECON PICs: An invalid private op‐
code was received from the NS486. The
message was ignored. This may happen
if the PIC firmware is out-of-date relative
to the SNE software.
Reload the OS and APP software for
the SNECON.
Replace the SNECON.
3762 ? Invalid Message on %4 SNECON PICs: An I2C message trans‐
action failed after the required number of
retries. A cause of this failure was a reply
message that had a valid checksum and
valid opcode but was invalid in any other
way. This may happen if the firmware of
the peripheral module is not working
properly or the SNECON software is too
old.
Reload the OS and APP software for
the SNECON. Replace the module be‐
ing addressed. Replace the SNECON.
3763 ! Arbitration Loss on %4 SNECON PICs: An I2C message trans‐
action failed after the required number of
retries and a cause of this failure was that
the PIC could not successfully arbitrate
its way to the bus. Other bus-master de‐
vices are using all the I2C bus bandwidth.
Contact Customer Support.
General Maintenance and Troubleshooting
4.6 Alarm Codes, Descriptions, and Suggested Actions
Maxum edition II Analyzer General Maintenance
Manual, August 2018, A5E42019842001 73

Table of Contents

Other manuals for Siemens Maxum II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Maxum II and is the answer not in the manual?

Siemens Maxum II Specifications

General IconGeneral
CategoryGas Chromatograph
Measurement PrincipleGas Chromatography
Communication ProtocolsModbus, Ethernet, Profibus
Operating Temperature-20°C to 55°C
Power Supply24 VDC or 100-240 VAC
Mountingrack mount
ApplicationsRefinery, chemical, petrochemical, natural gas
DetectorsThermal Conductivity Detector (TCD), Flame Ionization Detector (FID)
Sample InletsMultiple
DimensionsVaries by configuration (e.g., 19" rack mount)

Related product manuals