Mechanical systems 10 - 25
Siemens AG SPR8-220.061.03 Page 25 of 28 MOBILETT Plus HP
Medical Solutions Rev. 03 09.03 CS PS 24
Replacing the friction linings 10
Replacing the horizontal friction lining 10
Remove the joint covers.
Remove the locking nut.
Remove the screw and disc springs.
Replace the horizontal friction lining.
Remount the screw and disc springs (Fig. 27). Secure the screw with the locking nut.
Adjust the horizontal friction linings. Refer to the section: "Adjustment of the friction
linings", Page 10 - 4.
Reattach all covers.
Fig. 27
Locking nut
Disc springs
Friction lining