Siemens AG SPR8-220.061.03 Page 1 of 8 MOBILETT Plus HP
Medical Solutions Rev. 03 09.03 CS PS 24
Service programs 5
Entering service mode 5
• Remove the upper system cover.
To activate the service program, adjust the switch to the service position (S1:1 ->ON, on
board). This will display the program version:
Exit the service mode by adjusting the switch to the operating position (S1:1 -> OFF on
board D1). This will cause program execution to be aborted.
Selecting test programs with the kV+/- buttons 5
There are 16 different test programs available. The display shows the program number in
the kV window, the mAs display will be dark. Example, program 12:
Pressing the "kV+" button will display the next test program. When the last test program is
displayed, pressing the "kV+" button will show the program version again and so on.
Pressing the "kV-" button will display the previous test program. The lamp button has a
toggling function and is used both to start and stop the displayed test program. For exam-
ple, after the lamp button has been pressed, the formation of the capacitor bank starts
and the following is shown on the display:
The test program can be stopped by pressing the lamp button a second time. After the
lamp button is pressed, the display shows:
The program P12 has ended and a new service program can be selected with the kV+/-"
buttons or P12 can be started with the lamp button.
Error messages in service mode 5
Messages appearing in service mode, except in "Fast adaptation" are only displayed, they
are not logged into the memory history.
kV mAs
vEr x.x = Version x.x (depends on the software version installed)
kV mAs
P12
kV mAs
Ht tESt = High tension test
kV mAs
P12 StoP