Mechanical systems 10 - 3
Siemens AG SPR8-220.061.03 Page 3 of 28 MOBILETT Plus HP
Medical Solutions Rev. 03 09.03 CS PS 24
Counterbalance 10
• Remove the plastic plugs covering the friction linings in the column.
• Loosen the friction linings through the holes in the column (Fig. 4) and the lining in the
lower joint (Fig. 5).
• Check that the arm system stops in any position.
If does not stop, but moves lower or higher:
• Adjust the arm system with the adjustment screw (Fig. 6) until it balances in all positions.
If the arm moves lower, the screw must be tightened; if it moves up, the screw must be
loosened.
Fig. 4 For upper arm stability Fig. 5 For lower arm stability
Fig. 6
Friction lining
Friction lining
MOB00626
!
Adjustment screw