MOBILETT Plus HP SPR8-220.061.03 Page 8 of 10 Siemens AG
Rev. 03 09.03 CS PS 24 Medical Solutions
9 - 8 Replacement and repairs on the collimator
Light field relative to the X-ray field 9
Load a 35 x 35 cm (14" x 14") cassette with film.
Place the cassette on a table or a similar surface.
Set a SID of 100 cm (39"). Use a measuring tape.
Switch the main switch to position "B".
Switch on the collimator lamp and align the cassette. Collimate the light field to 25 x 25
cm (10" x 10").
Mark the four corners with coins and mark one of the corners with two coins (Fig. 13).
Set 52 kV, 2.0 mAs on the control panel.
Release an exposure.
Develop the film and measure the deviation between the light and the radiation field.
Acceptable deviation ± 1.7% of SID (Fig. 14).
Equipment required: Cassette, X-ray film and coins
If the MOBILETT is equipped with a DAP measuring system,
remove the DAP ionization chamber first.
X-ray radiation!
If not observed, risk of radiation exposure.
Use radiation protection!
Fig. 15
NOTE
CAUTION
MOB00650
E
E
E
E
Rotary flange