Replacement and repairs on the collimator 9 - 9
Siemens AG SPR8-220.061.03 Page 9 of 10 MOBILETT Plus HP
Medical Solutions Rev. 03 09.03 CS PS 24
• If the deviation is more than 1.7%, loosen the Allen screws marked "E" slightly and move
the rotary-flange to the desired position.
• Tighten the screws marked "E" (Fig. 15).
• Repeat the test and adjustment of the rotary flange position until the deviation between
the light and radiation fields is within the tolerance specified.
• Reinstall covers "A" and "B".