Replacement and repairs on the collimator 9 - 9
Siemens AG SPR8-220.061.03 Page 9 of 10 MOBILETT Plus HP
Medical Solutions Rev. 03 09.03 CS PS 24
If the deviation is more than 1.7%, loosen the Allen screws marked "E" slightly and move
the rotary-flange to the desired position.
Tighten the screws marked "E" (Fig. 15).
Repeat the test and adjustment of the rotary flange position until the deviation between
the light and radiation fields is within the tolerance specified.
Reinstall covers "A" and "B".