9 - 4 Replacement and repairs on the collimator
Replacing the collimator 9
• Switch off the unit and disconnect the power plug.
• Turn the single tank so that the collimator is facing upwards (Fig. 7).
• Secure the arm and the single tank with a rope (Fig. 7).
• Remove covers "A" and "B".
• Remove collimator cover "C".
Fig. 7
Equipment required: rope
C
A
B