Mechanical systems 10 - 17
Siemens AG SPR8-220.061.03 Page 17 of 28 MOBILETT Plus HP
Medical Solutions Rev. 03 09.03 CS PS 24
Replacement of the support arm cable harness 10
Position the unit with the single tank resting on a table. Secure the support arm with a
piece of rope (Fig. 19).
Remove covers "A" and "B".
Remove collimator cover "C".
Fig. 19
High Voltage!
Life-threatening hazard of electric shock. Never work with the
system open if the batteries are still connected and the capacitor
is charged. If the batteries are connected, the complete system is
powered on!
The capacitors in M10 and M11 may still be charged even if the
system is switched OFF and the mains cable is
disconnected. The capacitors in M10 and M11 must be consid-
ered charged until the protective measures listed in this section
have been performed.
Equipment required: Rope
MOB00439
C
A
B
DANGER