MOBILETT Plus HP SPR8-220.061.03 Page 16 of 28 Siemens AG
Rev. 03 09.03 CS PS 24 Medical Solutions
10 - 16 Mechanical systems
• Switch on the collimator lamp and check compliance of the light and radiation fields
according to the specifications in the section on "Replacement of collimator".
• Reattach all covers on the single tank.
• Adjust the adjustment screw so that the single tank is balanced. If any problems occur
(see section "Adjustment of arm system").
• Use the service program "P16 - Show the exposure counter" and record the values of the
exposure counter in the service protocol and in the unit documentation.
• Use the service program "P10 - Fast adaptation" to adjust the filament values to the new
single tank.
• Reattach the remaining covers.