Miscellaneous 8 - 3
Siemens AG SPR8-220.061.03 Page 3 of 8 MOBILETT Plus HP
Medical Solutions Rev. 03 09.03 CS PS 24
Replacement of mains cable/cable winder 8
Remove the side covers and the cover with the cassette compartment.
Remove the four (two on each side) screws that fixes the cable winder assembly
(pos. 1 / Fig. 2).
Pull out the cable winder assembly by approximately 10 cm.
Disconnect the main switch element by turning the red lock counterclockwise
(pos. 1 / Fig. 3).
Disconnect the wires at U9/U10 and ground connection (pos. 2 / Fig. 3).
Fig. 2 Fig. 3
Sharp edge!
The brake handle could be damaged.
Lock the brake handle, otherwise it will be damaged by the sharp
upper edge of the cable winder assembly.
MOB00624
1
2
MOB00618
1
CAUTION