MOBILETT Plus HP SPR8-220.061.03 Page 4 of 8 Siemens AG
Rev. 03 09.03 CS PS 24 Medical Solutions
8 - 4 Miscellaneous
• Remove the three cable ties (pos. 1 / Fig. 4).
• Remove the cable winder assembly.
• Remove the two cable ties (pos. 1 / Fig. 5).
• Remove the four screws and four nuts (pos. 2 / Fig. 5) and Y-plate (pos. 3 / Fig. 5).
• Disconnect the cable from upper plate (pos. 1 / Fig. 6).
• Remove the upper plate and distance ring (pos. 2 / Fig. 6).
• Loosen the brake carefully to loosen the tension of the cable winder spring.
• Remove the securing screw (pos. 1 / Fig. 8)
• Pull out the cable winder.