Miscellaneous 8 - 5
Siemens AG SPR8-220.061.03 Page 5 of 8 MOBILETT Plus HP
Medical Solutions Rev. 03 09.03 CS PS 24
• Clean cable winder box.
• Remove the plug, cable stop and ring (pos. 4 / Fig. 5).
If the cable winder has to be replaced:
• go to "B".
If the mains cable has to be replaced:
• Remove the four screws (pos. 3 / Fig. 6).
• Remove the connection plate and disconnect the mains cable (pos. 1 / Fig. 7).
• Install the new mains cable and connection plate.
"B"
• Reinstall the ring, cable stop and plug.
• Reinstall the cable winder.
• Reinstall the securing screw (pos. 1 / Fig. 8).
Fig. 7 Fig. 8
Wind the mains cable to the cable winder without any play.
1
MOB00622
1
NOTE