Mechanical systems 10 - 11
Siemens AG SPR8-220.061.03 Page 11 of 28 MOBILETT Plus HP
Medical Solutions Rev. 03 09.03 CS PS 24
Adjustment and replacement of the turn plate 10
The turn plate mechanism must be removed for adjustment or replacement.
Ensure compliance with the safety information in the section:
"Safety information and protective measures".
Remove the supporting wheels.
Jack up and block the unit securely 5 cm above the floor. This must be done properly so
that the unit does not tip over.
Remove battery charger D201 and motor power unit D102 (see section "7" and "11").
Disconnect all cables, ground connection and K68 according to (Fig. 13).
High Voltage!
Life-threatening hazard of electric shock. Never work with the
system open if the batteries are still connected and the capacitor
is charged. If the batteries are connected, the complete system is
powered on!
The capacitors in M10 and M11 may still be charged even if the
system is switched OFF and the mains cable is
disconnected. The capacitors in M10 and M11 must be consid-
ered charged until the protective measures listed in this section
have been performed.
Equipment required: 2 pieces of wood (50 x 70 x 500 mm)
Fig. 13
DANGER
Shaft
K68
Locking
Locking ring
ring