MOBILETT Plus HP SPR8-220.061.03 Page 12 of 22 Siemens AG
Rev. 03 09.03 CS PS 24 Medical Solutions
2 - 12 Display messages
Err 23 During exposure The HT inverter has
been interrupted
several times, e.g. by
overcurrent.
When the HT inverter has been inter-
rupted more than 3 times, the expo-
sure will be interrupted.
Faulty inverter/inverter control
Check the inverter current (see
sections "Error diagnosis HT and
single tank", "P12 - Inverter and
single tank test" and "P13 - Inverter
test").
The X-ray tube may have arced
during the exposure. Tube arcs can
occur if the first exposure after a
long interval is an exposure with
high kV values.
The unit has to be started with
several exposures at high mAs
values. The kV values selected
should be increased from 40 kV to
133 kV.
See also section "Error diagnosis
HT and single tank".
Dis-
played
messages
Mode of
operation
Cause Possible reasons and advice
* See section "Monitoring the kV and mA values"
** Wiring diagram X037E
Tab. 1

Table of Contents

Other manuals for Siemens MOBILETT Plus HP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens MOBILETT Plus HP and is the answer not in the manual?

Siemens MOBILETT Plus HP Specifications

General IconGeneral
BrandSiemens
ModelMOBILETT Plus HP
CategoryMedical Equipment
LanguageEnglish

Related product manuals