Display messages 2 - 21
Siemens AG SPR8-220.061.03 Page 21 of 22 MOBILETT Plus HP
Medical Solutions Rev. 03 09.03 CS PS 24
List of displayed messages
Err 3 Overvoltage > 142 kV
Err 4 Poor kV regulation
Err 5 No tube current
Err 6 Anode speed too slow
CAL 7 Exposure time too long
___ ___
Err 9 High-tension asymmetry
USE 10 Single tank > 50° or < 8°C, measured at the location of the temperature sensor
___ ___
CAL 12 Values in CMOS memory lost
CAL 13 Filament voltage, 60 V DC on D4 not present
CAL 14 Overvoltage in the capacitor bank (> 360 V)
USE 15 Preparation level held by the user > 20 s
Err 16 Capacitor bank charging too fast
Err 17 Capacitor bank charging too slow
Err 18 Capacitor voltage charging too slowly
CAL 19 Leakage current too high in the capacitor
USE 20 Radiation release button was released during the exposure
Err 21 Filament current incorrect for standby
Err 22 Filament current incorrect for exposure
Err 23 During exposure, the main inverter was interrupted > 3 times
USE 24 The selected kV and mAs values are lost
USE 25 The calculated anode load is lost
CAL 26 Error in RAM
CAL 27 Error in ROM
CAL 28 Error in the operating system
CAL 29 No single tank connected or an error in the connection
USE 30 The customer values stored in CMOS memory are lost
CAL 31 Replace backup battery < 2.8 V.
Err 32 Capacitance of the capacitor bank is too low
Err 34 Error log overfull
Err 35 Watchdog responded
USE 39 Battery voltage is too low