20 General
MOBILETT Plus/ -E/ -M SPR8-215.831.04.03.02 Siemens AG
07.05 CS PS 24
Page 20 of 64
Medical Solutions
2.3 Are the accessories being used free of
damage?
2.1.1 Damage, check:
SIM DAP (optional)
2.4 Are the system’s radiation protection
devices installed and damage-free? (no
radiation protection for the building)
2.15 Checking the radiation indicator,
checks:
SIE Radiation indicator
SIE Audible signal
2.16 Checking the manual exposure inter-
rupt, checks:
SIE Message "USE20"
SIE 10 short beep signals
2.5 Are the required operator documents
complete, present and legible?
The operating instructions must be
present. Ask the client for further required
operator documents and check if the docu-
ments are complete, present and legible.
2.6 Are all warning labels in place and rec-
ognizable?
2.1 Visual inspection, check:
SIM Labels
2.7 Are the operating symbols, light indica-
tors and button labels OK?
2.5 Maintenance tests, check:
SIE Display test
2.20 Checking the kV and mAs displays,
checks:
SIE Visibility of displays
SIE Correct displays
3. Electrical checks n.a.
3.1 Measurement of the protective conduc-
tor resistance
2.17 Protective conductor measurement,
check:
SIE Measured value
3.2 Measurement of leakage current or
equivalent leakage current
n.a.
3.2.1 Leakage current n.a.
3.2.2 Equivalent leakage current 2.18 Equivalent leakage current measure-
ment, check:
SIE Measured value
3.3 Measurement of patient leakage cur-
rent or equivalent patient leakage current
n.a.
3.3.1 Patient leakage current measurement n.a.
Technical Safety Checks,
Test certificate - check
Maintenance instructions - check/
Comments