Siemens AG SPR8-215.831.04.03.02 MOBILETT Plus/ -E/ -M
07.05 CS PS 24
General 21
Page 21 of 64
Medical Solutions
3.3.2 Equivalent patient leakage current
measurement
n.a.
4. Mechanical checks n.a.
4.1 Are all wall, ceiling and floor mountings
firmly attached, in place, and damage-free?
n.a.
4.2 Are all mechanically movable system
parts clean and running smoothly (lubri-
cated, if required)?
2.2.6 Transport safeguard, checks:
SIM Damage
SIM Locking
2.2.9 Arm system and single tank, checks:
SIM Movement of the arm system and the
single tank
SIM Movement of the single tank fork
SIM Single tank movement
4.3 Are the wire cables, chains, belts and
spindles free of wear and tear?
2.2.1 Rear wheels, check:
SIM Drive belt
4.4 Mobile equipment: Are the wheels, cas-
tors and brakes OK?
2.2.1 Rear wheels, checks:
SIM Brake lining
SIM Secure mounting
SIM Ease of movement
2.2.2 Front wheels, checks:
SIM Secure mounting
SIM Ease of movement
2.2.3 Supporting wheels, checks:
SIM Ease of movement
SIM Damage
2.2.5 Brakes, checks:
SIM Uniformity SIM Braking force
SIM Locking
4.5 Is there any unusual noise during oper-
ation (e.g. gearing)?
2.2.9 Arm system and single tank, check:
SIM Movement of the arm system and the
single tank
5. Functional checks n.a.
5.1 Function of the emergency-stop switch n.a.
Technical Safety Checks,
Test certificate - check
Maintenance instructions - check/
Comments