34 Inspection and maintenance
MOBILETT Plus/ -E/ -M SPR8-215.831.04.03.02 Siemens AG
07.05 CS PS 24
Page 34 of 64
Medical Solutions
SIM Screws and lock nuts
Check that all the screws and lock nuts on the arm system are tightened securely
(Fig. 13 / p. 33).
2.2.11.1 Adjustment screw
WARNING
If the screw is loosened, the arm might fall!
¹ When checking the adjustment screw, do not loosen the
nuts or turn the adjustment screw.
¹ Avoid loosening the screw and the nuts.
SIM Lock nuts
Check the lock nuts on the adjustment screw (Fig. 14 / p. 34).
SIM Base mounting
Check the base mounting in the chassis (Fig. 14 / p. 34).
2.2.11.2 Arm connecting piece
SIM Screws and nuts
Remove the cover of the lower connecting piece.
Remove the arm covers as well as the upper and lower cover plates of the column.
The lower column cover can be lifted up.
Fig. 14:
Pos. 1 Adjustment screw (check without torque)
Pos. 2 Foot
Fig. 15:
Pos. 1 Check without torque