Parameter lists
A.3 Diagnostic
SITRANS FUS080 transmitter
76 Operating Instructions, 07/2012, A5E03059912-02
Modbus
Address
Parameter Default
value
Value
range
Access
level
Description
40318 249 Pulse B overload fault
disappears
e.g. 0 0 ... 65535 Read Last time the fault disappeared.
Valid only if alarm is enabled
11 250 Consumption alarm
output enable
Yes No, Yes R/W Select 'Yes' to enable active alarm on
alarm output / call-up
40321 251 Consumption fault hours e.g. 0 h 0 ... 65535 Read Total hours fault active.
Valid only if alarm is enabled
40322 252 Consumption fault
counter
e.g. 0 0 ... 65535 Read Total number of faults.
Valid only if alarm is enabled
40323 253 Consumption fault
appears
e.g. 0 h 0 ... 65535 Read First time the fault appeared.
Valid only if alarm is enabled
40326 254 Consumption fault
disappears
e.g. 0 0 ... 65535 Read Last time the fault disappeared.
Valid only if alarm is enabled
12* 255 Leakage alarm output
enable
Yes No, Yes R/W Select 'Yes' to enable active alarm on
alarm output / call-up
40329 256 Leakage fault hours e.g. 0 h 0 ... 65535 Read Total hours fault active.
Valid only if alarm is enabled
40330 257 Leakage fault counter e.g. 0 0 ... 65535 Read Total number of faults.
Valid only if alarm is enabled
40331 258 Leakage fault appears e.g. 0 h 0 ... 65535 Read First time the fault appeared.
Valid only if alarm is enabled
40334 259 Leakage fault
disappears
e.g. 0 0 ... 65535 Read Last time the fault disappeared.
Valid only if alarm is enabled
13* 260 Empty pipe alarm output
enable
Yes No, Yes R/W Select 'Yes' to enable active alarm on
alarm output / call-up
40337 261 Empty pipe fault timer e.g. 0 h 0 ... 65535 Read Total hours fault active.
Valid only if alarm is enabled
40338 262 Empty pipe fault counter e.g. 0 0 ... 65535 Read Total number of faults.
Valid only if alarm is enabled
40339 263 Empty pipe fault appears e.g. 0 h 0 ... 65535 Read First time the fault appeared.
Valid only if alarm is enabled
40342 264 Empty pipe fault
disappears
e.g. 0 0 ... 65535 Read Last time the fault disappeared.
Valid only if alarm is enabled
Service
17 510 Fixed flow mode enable Yes No, Yes R/W Select 'Yes' to force the device to show
a fixed flow value
Default must be "No"
Always manually reset the value to
"No"
40801 511 Fixed flow value 0 -1E+09 ...
1E+09
R/W Fixed flow value for enabled fixed flow