Commands
SIWAREX FTA
Note
If the module group is in operating status “Hardware fault” or operating error then
only commands (8) Load Default Values, (1) Service On, (2) Service Off and (9)
Acknowledge Error, (108) ABORT are accepted. All other commands are rejected
with data/operating error 21.
6.2 Command list
in following
operating
Service and adjustment commands
The SIWAREX must be switched to service
operation to perform the adjustment. A non-
adjusted scale cannot get out of service operation.
cycle
After the adjustment, service operation can be
switched off. Only then can the scale accept
weighing commands.
operation
The beginning of the characteristic curve - zero
point of the scale - is defined with the momentary
dead-load.
operation
Adjustment weight 1 valid
The first adjustment weight is assigned with the
momentary weight.
operation
Adjustment weight 2 valid
The second adjustment weight is assigned with the
momentary weight.
operation
Adjustment weight 3 valid
The third adjustment weight is assigned with the
momentary weight.
operation
Adjustment weight 4 valid
The fourth adjustment weight is assigned with the
momentary weight.
operation