Messages and Diagnostics
SIWAREX FTA
7-143
Error
No.
Data and
command errors -
description
Solution
58
Log text contains
angle-brackets
Log text contains angle-brackets. Remove the angle-brackets
from the log text.
59
Reserved
60
Reserved
61
Reserved
62-
75
Log or MMC
command not
possible
62
Logging not
possible since the
weight lies outside
of the weighing
range
The logs can only be output within the weighing range in
calibratable operation (Code OIML in DR3).
63
Logging output not
possible since the
SIWATOOL FTA
driver is active
The log cannot be output at this time since the SIWATOOL FTA
program is active on the interface. Disconnect the PC and
connect the printer.
64
Logging output not
possible since the
printer is not ready
The log cannot be printed since the printer is not ready. Check
the printer.
65
Logging output not
possible since a
print job is still
running
The log cannot be printed since another print job is still running.
Wait until the current print job is complete.
66
Logging not
possible since the
MMC is full
Logging is not possible since the MMC is full. Switch MMCs.
67
Logging task
cannot be
executed, since
MMC is defective or
non-existent
Logging is not possible since the MMC is defective. Switch
MMCs.
68
Log output not
possible since there
is no data available.
Repeating the last log printout not possible since there was no
print-out or the stored data is defective.
69
Reserved
70
Logging output not
possible since the
log ID is not in the
print text
Log output is not possible since the content of the text does not
contain the log ID. Define the log ID in the text. This entry is
necessary if the calibratable weight is contained in the log text in
calibratable operation (Specifications Code OIML in DR 3).
71
Logging output not
possible since there
was noch change of
the measured value
After logging status standstill must have been inactive, otherwise
a further logging is rejected.
72
Reserved
73
Reserved
74
Reserved
75
Reserved
76-
93
Plausibility error
Calibration