64 57 498 D3543
100 D3543.201.01.05.02 08.2016
4Operation Sirona Dental Systems GmbH
4.5Dentist element Operating Instructions INTEGO
4.5.8.2
Setting the speed on the EasyPad
Speeds are stored on the favorite keys 1, 2, and 3 that can be accessed
by pressing them. Individual speeds can be saved for each user profile on
the favorite keys.
On the C
+
electronic foot switch, spray activation is saved on the favorite
keys. The settings can also be made via the 4-way switch plate.
Accessing saved speed
✔ The electric motor is removed from the holder.
✔ The set speed is shown on the EasyPad display.
➢ Press one of the favorite keys briefly (<1 s).
The speed saved on the favorite key is shown in rpm (revolutions
per minute) on the EasyPad display.
By moving the 4-way switch plate of the C
+
electronic foot switch to the
right, the settings of the favorite keys can be accessed consecutively.
Changing the speed
➢ Press and hold the favorite key 1 or 3 (> 1 s).
The speed is increased or reduced.
The speed can also be increased or reduced by moving the 4-way switch
plate of the C
+
electronic foot switch up or down.
Electric motor, increments with favorite keys
IMPORTANT
Increments
The size of the increments depends on the speed range setting.
From 100 to 1,000 rpm = 100 rpm increments (with eControl option)
From 1,000 to 2,000 rpm = 200 rpm increments
From 2,000 to 5,000 rpm = 500 rpm increments
From 5,000 to 10,000 rpm = 1,000 rpm increments
From 10,000 to 20,000 rpm = 2,000 rpm increments
From 20,000 to 40,000 rpm = 5,000 rpm increments
Please note that the speed of the burr depends on the selected straight
or contra-angle handpiece.