Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.5Components of the water unit
båÖäáëÜ
5.5.5 Change the water and air filters
If you notice any changes in media flows, check the water and air filters
for permeability. The filters must be changed, if necessary.
with treatment centers with Turn option
With treatment centers without the Ambidextrous option, the water and air
filters are located in the water unit. With treatment centers with the
Ambidextrous option, the water and air filters are located in the base of
the chair.
To reorder the filters, see "Spare parts and consumables" [ → 299].
1. Turn the treatment center OFF at the mains switch.
The water and air supply are switched off.
2. With treatment centers without the Ambidextrous option: open the
maintenance flap A of the water unit.
With treatment centers with the Ambidextrous option: convert the
water unit to the right side so that the filters are accessible,
seeConverting the Turn water unit for right-handers / light-handers
[ → 161].
3. Any remaining water flows out of the water filter when it is opened.
You should therefore place an absorbent cloth underneath the filter.
Then unscrew and remove the screw-on cap of the water filter B and/
or the air filter C.
4. Check the filters and replace them if necessary.