Sirona Intego Pro - Clean and Disinfect;Sterilize the Components of the

Sirona Intego Pro
326 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
64 57 498 D3543
D3543.201.01.05.02 08.2016
231
Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.3Instruments and instrument hoses
båÖäáëÜ
5.3.5.3
Clean and disinfect/sterilize the components of the ApexLocator
Care, cleaning, and disinfecting agents
Disconnect the mucosal electrode from the connection cable. The metal
hook can be sterilized, and the connection cable can be disinfected by
wiping.
The file clamp for manual measurement can be sterilized with the
connection cable.
The apex adapter and its connection cable can be disinfected by wiping.
ApexLocator, no disinfectant between contacts
The silicone isolation sleeve is a disposable item. It must be changed
after each patient. The silicone isolation sleeve must be sterilized prior to
initial use.
To reorder the silicone isolation sleeve, see "Spare parts and
consumables" [ 299].
Sterilize the root canal files according to the manufacturer's instructions.
NOTICE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning and disinfecting agents approved by Sirona, see
"Care, cleaning, and disinfecting agents" [ 205].
134°C 274°F
3 min
134°C 274°F
3 min
IMPORTANT
To guarantee conductivity, no disinfectant must be allowed to penetrate
the electrical contacts.
2
134°C
134°C 274°F
3 min
0
1
2
3
s
i
r
o
n
a
.
B
L
I
S
O
C