Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.3Instruments and instrument hoses
båÖäáëÜ
5.3.2.2
Purging function on the EasyTouch
Opening the purging dialog
✔ All instruments are in their holders.
✔ The
Start dialog
is displayed on the touchscreen.
1. Press the
Sub-dialog
fixed key.
The
Start
sub-dialog is displayed.
2. Touch the
Purge function
key.
Error message: Deposit instruments
If the
Deposit instruments
display appears after the purge function has
been started, the treatment center has detected that not all of the
instruments have been placed in their holders.
➢ Check the instruments marked with a warning triangle on the
touchscreen.
When all of the instruments have been put in the holders, the
purge program will continue.
Error message: Refill water (only for stand-alone water supply)
If the
Refill water
display appears after the purge function has been
started, there is not sufficient water in the disinfectant tank of the water
unit to purge the water lines. The purge function cannot be started with
insufficient water.
➢ Mix distilled water with the DENTOSEPT P disinfectant at a ratio of
100:1 (1 liter of water, 10 ml DENTOSEPT P) and fill this into the
disinfectant tank of the water unit. For more information, please refer
to the section "Stand-alone water supply for the Comfort water unit".
When sufficient water has been refilled, the purge program
continues.
Setting the purge time and starting the purge function
The purge time of the removed instruments can be set between 20 and
120 seconds.
Setup
San
Aqua
Apex Dist
Auto