64 57 498 D3543
216 D3543.201.01.05.02 08.2016
5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.3Instruments and instrument hoses Operating Instructions INTEGO
1. Remove the instruments to be purged from the holder and hold them
over the cuspidor or over a watertight container with sufficient
capacity. With the CS dentist element: Move the swivel arm of the
instruments to be purged into the operating position so that the weight
of the instrument hoses allows them to remain in this position.
After an instrument is removed, the purge process starts
automatically after about 5 seconds (except for the 3-way
syringe). The removed instruments are purged with water for the
duration of the set purging time.
2. Place the removed instruments back in their holders after the purging
process.
3. Hold the 3-way syringe of the dentist and assistant element over the
cuspidor or over a watertight container with sufficient capacity and
press the water key for at least 20 seconds.
4. Press the favorite key
2 / Setup
to end the purge program.
The time is displayed on the EasyPad display.
The treatment center is again ready for operation.
Interrupting the purging process
➢ Press the foot pedal to interrupt the purging process. If the instrument
is not put back in the holder, the purging process can be continued by
pressing the foot pedal again up to the end of the set purging time.
Canceling the purge function
The purging function can be stopped at any time.
➢ Press the
Counterclockwise rotation/user profile
fixed key.
The time is displayed on the EasyPad display.
CAUTION
Do not purge more than two burs simultaneously.
If the treatment center features the option of a third burr drive, the water
flow rate is reduced when purging simultaneously.
➢ Do not remove more than two burr drives for purging. Purge the third
burr drive separately afterwards.